In a cannabis extract, the major cannabinoids can be present at concentrations far above the minor compounds. One dilution can bring an abundant analyte into the instrument’s calibration range while pushing a minor analyte below the level the method can quantify.

What the validation measured

McRae and Melanson coupled LC–MS/MS with a broad calibration range and multiple dilutions of plant extracts. They reported a quantifiable range of 0.002 to 200 mg/g in plant material for the method. They also evaluated performance in a candidate cannabis reference material and measured precision across seven cannabis samples.

How to read an individual lab result

Ask which dilution supplied the reported result for each analyte and whether its instrument response fell within calibration. A minor compound that falls below the quantification limit in a highly diluted extract may still be quantifiable at a lower dilution. The method’s coverage depends on both the calibration and the dilution used.

Source

McRae and Melanson, Analytical and Bioanalytical Chemistry · PubMed record